One-to-one debugging skills for high-speed flat mask machine

Time:0000-00-00 00:00:00 Author:Suny Group

1. Turn on the start-stop switch and the power switch, and the built-in light of the power switch is on.

2. Adjust the air pressure. The air pressure valve is under the workbench. Open the door of the equipment and you can see the air pressure valve. Adjust the pressure according to the actual situation. When some mechanisms are not synchronized, increase the pressure. The working air pressure is generally Adjust within the range of 0.2~0.6MPa. 3. When the equipment is stopped, turn on the power switch of the ultrasonic generator and press the "Sound Wave Test" switch. If the amplitude indicator indicator rises sharply and exceeds 100, it means that the frequency is too far from the resonance point and needs to be adjusted. "Sound wave adjustment" knob. Look at the amplitude display, press the "Sound Wave Test" check switch with your left hand (not more than 3 seconds), and turn the "Sound Wave Adjustment" knob with your right hand to adjust the pointer of the amplitude display to the lowest point. When adjusting, if the amplitude indicator indicator does not fall but rises, turn the "sound wave adjustment" knob in the opposite direction. The reading of the amplitude indicator indicator is within 50, and the reading of the ammeter indicator is below 0.6A, which is the best working condition. After adjusting the sound wave, tighten the butterfly nut and close the protective cover.

Fully Automatic Mask Making Machine 1+1

Fourth, the cooling fan inspection, check whether the cooling fan is working normally, to ensure that the ultrasonic transducer can work for a long time when it is working.

Five, check the pneumatic parts

(1) Turn on the power, connect the gas path, and switch the manual and automatic switch to manual gear.

(2) Turn on the material pan switch, check whether the material pan can run quickly and steadily and locate accurately, and turn off the material pan switch after checking.

(3) Turn on the mouth clamp switch to check whether the mask clamp is operating normally, and turn off the mask clamp switch after checking.

(4) Turn on the collecting cylinder switch to check whether the collecting cylinder is operating normally, and turn off the collecting cylinder switch after checking.

(5) Turn on the sideband fusion switch to see if the sideband die can be pressed down normally.

(6) Turn on the right switch of the ear strap to see if the rotating cylinder on the right moves and moves in place.

(7) Turn on the turn left switch of the ear strap to see if the rotating cylinder on the left moves and moves in place.

(8) Turn on the ear strap cutting switch and check whether the ear strap cutting cylinder is working and can reach the designated cutting position.

(9) Turn on the ear strap return switch, check whether the ear strap return small cylinder can work normally, and the ear strap clip can be loosened.

(10) Turn on the ear strap welding switch and check whether the ear strap pressure mold can work and reach the designated position.